MDT Scaling

To have any useful meaning, the counts from the transducer must be scaled to position units. The Position Scale and Position Offset parameters define the position units as a function of transducer counts. This topic describes how to correctly calculate these parameters for an MDT transducer.

Delta Recommends using the Scale/Offset Wizard for scaling the position. If you need to do it manually, read this topic.

Scaling Counts to Position Units

The RMC calculates the Actual Position every control-loop time using the following formula:

Actual Position [pu] = (Counts [cnt] x Position Scale [pu/cnt]) + Position Offset [pu]

 

Note:
If the Actual Position filter is applied, the RMC filters the Actual Position after calculating it with the above formula.

 

Manually Calculating the Scale and Offset

The Scale/Offset Wizards provide the easiest method of scaling your axis. If you prefer to do it manually, read this section.

Determining the Scale and Offset

There are two methods of finding the correct Position Scale and Position Offset.

ClosedMethod 1: P0/P1 Calculation

The accuracy of this method depends on how accurately you can measure two positions of the axis.

  1. Physically measure the axis position at two points and record the value of the counts register at each point. Call the smaller measured position P0, and its corresponding Counts C0. Call the greater measured position P1, and its corresponding Counts C1.

  2. Calculate the Position Scale with the following equation:

    Position Scale = (P0-P1)/(C0-C1)

  3. Calculate the Position Offset with the following equation:

    Position Offset = P0 - Position Scale x C0

ClosedMethod 2: Using the MDT Calibration Number

  1. Obtain the following information (usually from the transducer label or data sheet):

    • Calibration Number (calibration constant)  in μs/in or μs/μm (typically about 9 μs/in)

    • Number of Recirculations (if it has no recircs, use a value of 1)

  2. Calculate the Position Scale with the following formula:

RMC75 and RMC200:

RMC150:

where

  1. If you wish to reverse the direction of the feedback, make the Scale negative.

  2. Download the Scale to the RMC.

  3. Calculate the Position Offset:

  1. Download the Offset to the RMC.

Example

  1. Connected to an RMC75 is an MDT transducer with a calibration number of 9.1615 μs/in and no recirculations (this means it has just 1 recirculation). The system needs to be scaled to inches.

    • To calculate the Scale:

    • Download the Scale to the RMC.

    • To calculate the Offset: Move the axis to where it should be 0. The Actual Position shows 2.4528. The Position Offset should be: 0 - 2.4528 = -2.4528.

    • Download the Offset to the RMC.

 

  1. Connected to an RMC75 is an MDT transducer with a calibration number of 9.012 μs/in and 2 recirculations. The system needs to be scaled to inches.

    • To calculate the Scale:

    • Download the Scale to the RMC.

    • To calculate the Offset: Move the axis to where it should be 10.125. The Actual Position shows 11.241. The Position Offset should be: 10.125 - 11.241 = -1.116.

    • Download the Offset to the RMC.

 

  1. Connected to an RMC150 is an MDT transducer with a calibration number of 9.1313 μs/in and no recirculations (this means it has just 1 recirculation). The system needs to be scaled to millimeters.

    • We must first calculate the scale, but in order to do that we need to know how many position units there are in an inch: 25.4 mm/in.
       

    • To calculate the Scale:

    • Download the Scale to the RMC.

    • To calculate the Offset: Move the axis to where it should be 0 mm. The Actual Position shows 24.8. The Position Offset should be: 0 - 24.8 = -24.8.

    • Download the Offset to the RMC.

 

See Also

Scaling Overview | Scale/Offset Wizard | Position Scale | Position Offset


Send comments on this topic.

Copyright © 2024 Delta Computer Systems, Inc. dba Delta Motion